1. Intercomparison and characterization of 23 Aethalometers under laboratory and ambient air conditions : procedures and unit-to-unit variabilitiesAndrea Cuesta-Mosquera, Griša Močnik, Luka Drinovec, Thomas Müller, Sascha Pfeifer, Maria Cruz Minguillon, Björn Briel, Paul Buckley, Vadimas Dudoitis, Jesús Yus-Díez, 2021, original scientific article Abstract: Abstract. Aerosolized black carbon is monitored worldwide to quantify its impact on air quality and climate. Given its importance, measurements of black carbon mass concentrations must be conducted with instruments operating in quality-checked and ensured conditions to generate data which are reliable and comparable temporally and geographically. In this study, we report the results from the largest characterization and
intercomparison of filter-based absorption photometers, the Aethalometer model AE33, belonging to several European monitoring networks. Under controlled laboratory conditions, a total of 23 instruments measured mass
concentrations of black carbon from three well-characterized aerosol
sources: synthetic soot, nigrosin particles, and ambient air from the urban
background of Leipzig, Germany. The objective was to investigate the
individual performance of the instruments and their comparability; we
analyzed the response of the instruments to the different aerosol sources
and the impact caused by the use of obsolete filter materials and the
application of maintenance activities. Differences in the instrument-to-instrument variabilities from equivalent black carbon (eBC) concentrations reported at 880 nm were determined before maintenance
activities (for soot measurements, average deviation from total least square regression was −2.0 % and the range −16 % to 7 %; for nigrosin measurements, average deviation was 0.4 % and the range −15 % to 17 %), and after they were carried out (for soot measurements, average deviation was −1.0 % and the range −14 % to 8 %; for nigrosin measurements, the average deviation was 0.5 % and the range −12 % to 15 %). The deviations are in most of the cases explained by the type of filter material employed by the instruments, the total particle load on the filter, and the flow calibration. The results of this intercomparison activity show that relatively small
unit-to-unit variability of AE33-based particle light absorbing measurements is possible with well-maintained instruments. It is crucial to follow the guidelines for maintenance activities and the use of the proper filter tape in the AE33 to ensure high quality and comparable black carbon (BC) measurements among international observational networks. Keywords: Aethalometer AE33, filter photometers Published in RUNG: 29.11.2024; Views: 227; Downloads: 0 Full text (5,57 MB) This document has many files! More... |
2. Actual and perceived wood combustion pollution – The case of a rural mountainous areaKristina Glojek, Griša Močnik, Honey Dawn Alas, Andrea Cuesta-Mosquera, Luka Drinovec, Asta Gregorič, Kay Weinhold, Thomas Müller, Dejan Cigale, Martin Rigler, Dominik van Pinxteren, Maik Merkel, Matej Ogrin, Martina Ristorini, Hartmut Herrmann, Alfred Wiedensohler, 2022, published scientific conference contribution abstract Keywords: local air quality, aerosol measurements, perception, heating habits Published in RUNG: 27.02.2023; Views: 2413; Downloads: 0 This document has many files! More... |
3. Experimental dataset - A dual-wavelength photothermal aerosol absorption monitor: design, calibration and performanceLuka Drinovec, Uroš Jagodič, Luka Pirker, Miha Škarabot, Mario Kurtjak, Kristijan Vidović, Luca Ferrero, Bradley Visser, Jannis Röhrbein, Ernest Weingartner, Daniel Kalbermatter, Konstantina Vasilatou, Tobias Bühlmann, Celine Pascale, Thomas Müller, Alfred Wiedensohler, Griša Močnik, complete scientific database of research data Keywords: photothermal interferometer, aerosol light absorption coefficient Published in RUNG: 14.01.2022; Views: 2715; Downloads: 35 Full text (531,36 MB) This document has many files! More... |
4. Intercomparison and characterization of 23 Aethalometers under laboratory and ambient air conditions: : procedures and unit-to-unit variabilitiesAndrea Cuesta-Mosquera, Griša Močnik, Luka Drinovec, Thomas Müller, Sascha Pfeifer, Maria Cruz Minguillon, Björn Briel, Paul Buckley, Vadimas Dudoitis, Javier Fernández-García, 2021, original scientific article Abstract: Aerosolized black carbon is monitored worldwide to quantify its impact on air quality and climate. Given its importance, measurements of black carbon mass concentrations must be conducted with instruments operating in quality-checked and ensured conditions to generate data which are reliable and comparable temporally and geographically.
In this study, we report the results from the largest characterization and intercomparison of filter-based absorption photometers, the Aethalometer model AE33, belonging to several European monitoring networks. Under controlled laboratory conditions, a total of 23 instruments measured mass concentrations of black carbon from three well-characterized aerosol sources: synthetic soot, nigrosin particles, and ambient air from the urban background of Leipzig, Germany. The objective was to investigate the individual performance of the instruments and their comparability; we analyzed the response of the instruments to the different aerosol sources and the impact caused by the use of obsolete filter materials and the application of maintenance activities.
Differences in the instrument-to-instrument variabilities from equivalent black carbon (eBC) concentrations reported at 880 nm were determined before maintenance activities (for soot measurements, average deviation from total least square regression was −2.0 % and the range −16 % to 7 %; for nigrosin measurements, average deviation was 0.4 % and the range −15 % to 17 %), and after they were carried out (for soot measurements, average deviation was −1.0 % and the range −14 % to 8 %; for nigrosin measurements, the average deviation was 0.5 % and the range −12 % to 15 %). The deviations are in most of the cases explained by the type of filter material employed by the instruments, the total particle load on the filter, and the flow calibration.
The results of this intercomparison activity show that relatively small unit-to-unit variability of AE33-based particle light absorbing measurements is possible with well-maintained instruments. It is crucial to follow the guidelines for maintenance activities and the use of the proper filter tape in the AE33 to ensure high quality and comparable black carbon (BC) measurements among international observational networks. Keywords: black carbon, aerosol, absorption, filter absorption photometer, aethaloemter Published in RUNG: 03.05.2021; Views: 2854; Downloads: 0 This document has many files! More... |