Naslov: | Optimized frequency dependent photothermal beam deflection spectroscopy |
---|
Avtorji: | ID Franko, Mladen, univerza v novi gorici (Avtor) ID Korte, Dorota, univerza v novi gorici (Avtor) ID Cabrera, Humberto, Instituto Venezolano de Investigaciones Científicas (Avtor), et al. |
Datoteke: |
Gradivo nima datotek, ki so prostodostopne za javnost. Gradivo je morda fizično dosegljivo v knjižnici fakultete, zalogo lahko preverite v COBISS-u. |
---|
Jezik: | Angleški jezik |
---|
Vrsta gradiva: | Delo ni kategorizirano |
---|
Tipologija: | 1.01 - Izvirni znanstveni članek |
---|
Organizacija: | UNG - Univerza v Novi Gorici
|
---|
Opis: | In the paper the optimization of the experimental setup for photothermal beam deflection spectrometry is performed by analyzing the influence of its geometrical parameters (detector and sample position, probe beam radius and its waist position etc) on the detected signal. Furthermore,
the effects of fluid’s thermo-optical properties, for optimized geometrical configuration, on the
measurement sensitivity and uncertainty determination of sample thermal properties is also studied.
The examined sample is a recently developed CuFeInTe3 material. Results show, that it is a complex problem to choose the proper geometrical configuration as well as
sensing fluid to enhance the sensitivity of the method. A signal enhancement is observed at low modulation frequencies by placing the sample in acetonitrile (ACN), while at high modulation frequencies the sensitivity is higher for measurements made in air. For both, detection in air and
acetonitrile the determination of CuFeInTe3 thermal properties is performed. The determined values of thermal diffusivity and thermal conductivity are (0.048 ± 0.002) × 10−4 m2 s−1 and 4.6 ± 0.2 W m−1 K−1 and (0.056 ± 0.005) × 10−4 m2 s−1 and 4.8 ± 0.4 W m−1 K−1 for ACN and air,
respectively. It is seen, that the determined values agree well within the range of their measurement uncertainties for both cases, although the measurement uncertainty is two times lower for the measurements in ACN providing more accurate results. The analysis is performed by the use of
recently developed theoretical description based on the complex geometrical optics. It is also shown, how the presented work fits into the current status of photothermal beam deflection spectroscopy. |
---|
Ključne besede: | Optitermi, na spektrometrija, toplotna difuzivnost, toplotna prevodnost, polprevodniki, vpliv topila |
---|
Leto izida: | 2016 |
---|
Št. strani: | 12 |
---|
Številčenje: | 13, 12 |
---|
PID: | 20.500.12556/RUNG-2755-cde09edb-49ac-1e1c-618b-e8c7ccb974e1 |
---|
COBISS.SI-ID: | 4562939 |
---|
DOI: | http://dx.doi.org/10.1088/1612-2011/13/12/125701 |
---|
NUK URN: | URN:SI:UNG:REP:0TOUYOCI |
---|
Datum objave v RUNG: | 07.11.2016 |
---|
Število ogledov: | 5757 |
---|
Število prenosov: | 0 |
---|
Metapodatki: | |
---|
:
|
Kopiraj citat |
---|
| | | Skupna ocena: | (0 glasov) |
---|
Vaša ocena: | Ocenjevanje je dovoljeno samo prijavljenim uporabnikom. |
---|
Objavi na: | |
---|
Postavite miškin kazalec na naslov za izpis povzetka. Klik na naslov izpiše
podrobnosti ali sproži prenos. |