Naslov: | The new instrument using a TC–BC (total carbon–black carbon) method for the online measurement of carbonaceous aerosols |
---|
Avtorji: | ID Rigler, Martin, Aerosol d.o.o. (Avtor) ID Drinovec, Luka, Institut Jozef Stefan (Avtor) ID Lavrič, Gašper, Aerosol d.o.o. (Avtor) ID Vlachou, Anastasia, Paul Scherrer Institut (Avtor) ID Prévôt, André S. H., Paul Scherrer Institut (Avtor) ID Jaffrezo, Jean-Luc, IGE CNRS (Avtor) ID Stavroulas, Iasonas, Cyprus Institute (Avtor) ID Sciare, Jean, Cyprus Institute (Avtor) ID Burger, Judita, Agencija RS za okolje (Avtor) ID Krajnc, Irena, Agencija RS za okolje (Avtor) ID Turšič, Janja, Agencija RS za okolje (Avtor) ID Hansen, Anthony D. A., Magee Scientific Corp. (Avtor) ID Močnik, Griša, Institut Jožef Stefan (Avtor) |
Datoteke: | Rigler_-_AMT_2020_-_Suppl_-_The_new_instrument_using_a_TC–BC_total_carbon–black_carbon_method_for_the_online_measurement_of_carbonaceous_aerosols.pdf (226,45 KB) MD5: 2CE41B01E67C915ADEE07567F8604B92
|
---|
Jezik: | Slovenski jezik |
---|
Vrsta gradiva: | Delo ni kategorizirano |
---|
Tipologija: | 1.01 - Izvirni znanstveni članek |
---|
Organizacija: | UNG - Univerza v Novi Gorici
|
---|
Opis: | We present a newly developed total carbon analyzer (TCA08) and a method for online speciation of carbonaceous aerosol with a high time resolution. The total carbon content is determined by flash heating of a sample collected on a quartz-fiber filter with a time base between 20 min and 24 h. The limit of detection is approximately 0.3 µg C, which corresponds to a concentration of 0.3 µg C m−3 at a sample flow rate of 16.7 L min−1 and a 1 h sampling time base. The concentration of particulate equivalent organic carbon (OC) is determined by subtracting black carbon concentration, concurrently measured optically by an Aethalometer®, from the total carbon concentration measured by the TCA08. The combination of the TCA08 and Aethalometer (AE33) is an easy-to-deploy and low-maintenance continuous measurement technique for the high-time-resolution determination of equivalent organic and elemental carbon (EC) in different particulate matter size fractions, which avoids pyrolytic correction and the need for high-purity compressed gases. The performance of this online method relative to the standardized off-line thermo-optical OC–EC method and respective instruments was evaluated during a winter field campaign at an urban background location in Ljubljana, Slovenia. The organic-matter-to-organic-carbon ratio obtained from the comparison with an aerosol chemical speciation monitor (ACSM) was OM/OC=1.8, in the expected range. |
---|
Ključne besede: | total carbon, aeroosl, black carbon, carbonaceous matter |
---|
Verzija publikacije: | Objavljena publikacija |
---|
Leto izida: | 2020 |
---|
Št. strani: | 4333-4351 |
---|
Številčenje: | 2020 |
---|
PID: | 20.500.12556/RUNG-5689 |
---|
COBISS.SI-ID: | 25335555 |
---|
DOI: | 10.5194/amt-13-4333-2020 |
---|
NUK URN: | URN:SI:UNG:REP:IWRINC8T |
---|
Datum objave v RUNG: | 17.08.2020 |
---|
Število ogledov: | 3798 |
---|
Število prenosov: | 80 |
---|
Metapodatki: | |
---|
:
|
Kopiraj citat |
---|
| | | Skupna ocena: | (0 glasov) |
---|
Vaša ocena: | Ocenjevanje je dovoljeno samo prijavljenim uporabnikom. |
---|
Objavi na: | |
---|
Postavite miškin kazalec na naslov za izpis povzetka. Klik na naslov izpiše
podrobnosti ali sproži prenos. |