Abstract:There exists a lack of aerosol absorption measurement techniques with low uncertainties and without artefacts. We have developed the two-wavelength Photothermal Aerosol Absorption Monitor (PTAAM-2λ), which measures the aerosol absorption coefficient at 532 and 1064 nm. Here we describe its design, calibration and mode of operation and evaluate its applicability, limits and uncertainties. The 532 nm channel was calibrated with ∼ 1 µmol mol−1 NO2, whereas the 1064 nm channel was calibrated using measured size distribution spectra of nigrosin particles and a Mie calculation. Since the aerosolized nigrosin used for calibration was dry, we determined the imaginary part of the refractive index of nigrosin from the absorbance measurements on solid thin film samples. The obtained refractive index differed considerably from the one determined using aqueous nigrosin solution. PTAAM-2λ has no scattering artefact and features very low uncertainties: 4 % and 6 % for the absorption coefficient at 532 and 1064 nm, respectively, and 9 % for the absorption Ångström exponent. The artefact-free nature of the measurement method allowed us to investigate the artefacts of filter photometers. Both the Aethalometer AE33 and CLAP suffer from cross-sensitivity to scattering – this scattering artefact is most pronounced for particles smaller than 70 nm. We observed a strong dependence of the filter multiple scattering parameter on the particle size in the 100–500 nm range. The results from the winter ambient campaign in Ljubljana showed similar multiple scattering parameter values for ambient aerosols and laboratory experiments. The spectral dependence of this parameter resulted in AE33 reporting the absorption Ångström exponent for different soot samples with values biased 0.23–0.35 higher than the PTAAM-2λ measurement. Photothermal interferometry is a promising method for reference aerosol absorption measurements.
Keywords:aerosol absorption, calibration, black carbon
Year of publishing:2022
Number of pages:str. 3805–3825
Numbering:Vol. 15, iss. 12
PID:20.500.12556/RUNG-7397 New window
COBISS.SI-ID:99980547 New window
UDC:54
ISSN on article:1867-1381
DOI:10.5194/amt-2022-21 New window
NUK URN:URN:SI:UNG:REP:YAPXBUJG
Publication date in RUNG:28.06.2022
Views:1208
Downloads:25
Metadata:XML RDF-CHPDL DC-XML DC-RDF
:
Copy citation
  
Average score:(0 votes)
Your score:Voting is allowed only for logged in users.
Share:Bookmark and Share


Hover the mouse pointer over a document title to show the abstract or click on the title to get all document metadata.

Record is a part of a journal

Title:Atmospheric measurement techniques
Shortened title:Atmos. meas. tech.
Publisher:Copernicus Publications
ISSN:1867-1381
COBISS.SI-ID:517910809 New window

Document is financed by a project

Funder:ARRS - Slovenian Research Agency
Project number:P1-0385, P1-0099, I-0033

Secondary language

Language:Slovenian
Keywords:kemija, aerosoli, meritve, modeliranje, kalibriranje


Back